Symptoms |
Problem |
Solution |
1. The scanner does not power up. Sometimes you
would see one or two blinks of indicator (power) light after you press the
power button and then nothing. Sometimes you would also see brief flash of
the indicator light when you depress the power button. |
The power button does NOT latch in ON position. |
Replace
the power switch. |
Power supply issues. Particularly in LS-50 and LS-5000 units with serial
number starting with 4 (4XXXXX), this may be caused by
a bad IC in the inverter section, the part STRW6756. |
Check the voltages
Replace
the part STRW6756 manufactured by Sanken. |
|
|
|
2. Scanner does not complete initialization
properly. Power light blinks fast after the first movement sound, before
the second movement sound and before testing LED/CCD (you cannot see the LED
flashing red/green/blue).
See here. |
Autofocus assembly problem. Most likely
the proximity sensor in autofocus assembly is either dirty/obstructed or
dead. |
Disassemble the scanner remove optical scan head. Check if there is any
debris in the autofocus proximity sensor window. Clean if dirty. Replace if
that does not help. |
|
|
|
3. Scanner does not complete initialization
properly. Power light blinks fast after the second movement sound, before
testing LED/CCD (you do not see the LED flashing red/green/blue). A loud
whining sound is heard that would not stop until the scanner is turned
off. |
The limit switch (SHARP GP1A75E optical
interruptor) in the transport assembly is malfunctioning. |
Disassemble the scanner and check the wiring between the motherboard and
SHARP GP1A75E optical interrupter. If there are no problems with wiring -
replace the interrupter. |
|
|
|
4. Scanner does not complete
initialization properly. Power light blinks fast after the second movement sound, and after attempting testing
LED/CCD (you do not see the LED flashing red/green/blue).
See
here. |
Motherboard failure. |
Test voltage regulators on the motherboard.
There are 5 or 6 of them, and you can figure out the ratings from the part
numbers. Replace if failed. |
LED failure. |
Replace LED source. |
|
|
|
5. Scanner does not complete initialization
properly. Power light blinks fast after all motor tests are complete AND
after testing LED/CCD is complete (you can see the LED flashing
red/green/blue.) |
Something is malfunctioning in optical
light path.
- A
most likely cause is a dirty or misplaced imaging mirror. |
Check the imaging mirror.
Clean if dirty, position
properly. |
-
The LED is mis-positioned.
|
Adjust LED position. Move it forward
/backward in 1/32” increments and test again. |
- CCD
is mis-positioned |
Adjust CCD position. Move it up / down in
1/32” increments and test again. |
I have seen few cases where a flex cable between
the CCD assembly and the motherboard was ripped during unsuccessful repair.
It would result in the same symptoms. |
Repair or replace the flex cable. |
|
|
|
6. The LS-4000 scanner
passes power-up initialization fine (no fast blinking), but is not
recognized by a computer (in Windows - Device Manager does not see it), the
scan software reports that it cannot find a scanner.
The failure can also
manifest itself in two other different ways:
- intermittent loss of
communication with computer
- failure to boot up
properly |
There are two IC's on the motherboard,
that are directly responsible for IEEE communication: TSB41AB3 and
SAA7356HL. One or both may have failed.
Complete loss of communication is usually an indication of failed TSB41AB3.
Intermittent loss of communication is more
frequently an indication of failed SAA7356HL.
|
Replace TSB41AB3
and / or SAA7356HL. |
7. LS-4000:
I have seen few cases where a failed TSB41AB3
was pulling down the power supply voltages and the scanner was failing to
boot up properly. |
|
Replace TSB41AB3. |
8. The LS-4000 scanner
passes power-up initialization fine (no fast blinking), and is recognized by
a computer (in Windows - Device Manager), BUT - Nikon Scan does NOT see it. |
In
every case (that I saw), it was caused by a failed SAA7356HL. |
Replace SAA7356HL. |
|
|
|
9. The LS-50
(V) or LS-5000 scanner passes power-up initialization (no fast blinking),
but is not recognized by a computer (in Windows - Device Manager does not
see it), the scan software reports that it cannot find a scanner |
In most cases this indicative of damaged VBUSIN port in
EI-107 ASIC on the scanner's motherboard. |
Graeme Hardie
at
LincolnScan
proposed a work-around procedure. |
|
|
|
10.
The start-up without adapter inserted into a scanner completes properly.
However, when you insert MA-20/21 (or any other adapter), you will hear (at the
end of adapter initialization) a loud high-pitched whirring sound.
See this video.
If you power up the scanner
with the adapter inserted,
the scanner (main stepper motor) will make a loud high-pitched whirring sound at
the end of startup, see
here.
Sometimes the scanner will blink fast after the loud whirring sound, sometimes
not.
or
Image is shifted along the long axis
(image outside frame).
|
Both of this are indicative of the main stepper motor failure. For some
reason this failure is prevalent in late production number (SN4XXXXX) LS-50
and LS-5000.
If it is the second symptom (image
is shifted along the long axis),
this could also be caused by a slipping gears, see the cells below. |
Replace the main
stepper motor.
The replacement procedure is fairly straightforward, the problem is a
replacement motor. I could not find a suitable motor, so I have to scavenge
older (cheaper) models for these, |
|
|
|
11. Scanner stops working with SF-200 / SF-210 batch loader adapters.
When you insert SF-210 into the scanner - the power light blinks few times
but no motor sounds occur. The scanner reports failure to load film |
Possible failures: - Failed
motor in SF-200/SF-210
- Failed driver IC on the scanner motherboard |
Replace the motor in SF-200/SF-210 (original Minebea part PPN13KA11C -
good replacement is
PPN13PB11C2)
Replace
U119 on the motherboard:
NEC D16805 (Monolithic
H-Bridge Driver) |
|
|
|
12. Scanner makes loud grinding noises when SA-21 is
inserted, see here. |
- Failed driver IC on the scanner motherboard |
Replace U120
on the motherboard:
NEC A1603C (Stepper
Motor Driver) |
|
|
|
13. Image is shifted along the long axis
(image outside frame). |
Possible failures:
- main stepper motor
is bad
- the gear is slipping
on the main stepper motor shaft
- leadscrew gear slipping on the
leadscrew shaft. |
Disassemble the scanner remove optical scan head.
Replace the main
stepper motor.
Take the gear in question off the metal
shaft. Apply cyanoacrylate (superglue) and reattach it. |
|
|
|
14. Image is tilted. |
CCD is mis-positioned (angular) |
Adjust CCD angle. Move one side up / down
in 1/32” increments and test again |
|
|
|
15. Image is shifted along the short axis. |
CCD is in a wrong position |
Adjust CCD position. Move it left / right
in 1/32” increments and test again |
|
|
|
145. Image is stretched (has wrong aspect ratio). |
imaging lens is in a wrong position |
Adjust the position the imaging lens. Move it so
that the gap is 3.8mm, see
here. |
|
|
|
16. Image is doubled |
Imaging mirror was placed wrong side in. |
Flip the mirror to make sure is placed with correct side
facing in. See photo over set 8)
here. The mirror should be
sitting in the housing at an angle, reflective surface down (if the scanner
is placed upside down as in my instructions). |
|
|
|